DataSheet26.com

CLE-0922-400 PDF даташит

Спецификация CLE-0922-400 изготовлена ​​​​«Solidsense» и имеет функцию, называемую «Chlorine Sensor».

Детали детали

Номер произв CLE-0922-400
Описание Chlorine Sensor
Производители Solidsense
логотип Solidsense логотип 

2 Pages
scroll

No Preview Available !

CLE-0922-400 Даташит, Описание, Даташиты
Classic Line 4-Cl2-200 Sensor
Chlorine Sensor 0-200 ppm
Performance Characteristics
0BU
Part Number
CLE-0922-400
Nominal Range
0 to 200ppm
Sensitivity
Baseline20 oC
0.18 ± 0.05 µA/ppm
< ± 0.2 µA
Baseline Drift
(-20 to 40 oC)
0 to -0.3 ppm equivalent
Resolution
0.1 ppm
Response Time (T90)
Linearity
30 seconds
Linear
Long Term Output Drift < 2% signal/month
Operation Conditions
1BU
Temperature Range
-20 oC to 50 oC
Operating Humidity 15 to 90%RH non-condensing
Pressure Range
90 to 110 kPa
Bias Potential
0 mV
Storage Life
Storage Temperature
6 months in sealed container
0 oC to 20 oC
Expected Operating Life 2 years in air
Warranty
12 months from date of
despatch
Physical Characteristics
2BU
Weight
5 g (approx)
Orientation Sensitivity None
Note: PCB sockets are recommended for the sensor pin connection.
Soldering to the sensor should be avoided.
Note: All performance specifications are based upon the following environment
conditions: 20 oC, 50% relative humidity and 1 atm (1013 mBar or ambient
pressure)
Solidsense GmbH - Felix-Wankel-Str. 5 - 82152 Krailling, Germany
Tel: +49 89 893 255 21 – Fax: +49 89 850 9374 – [email protected] – www.solidsense.de
CLE 4-Cl2-200 rev120727.doc - Page 1 of 2









No Preview Available !

CLE-0922-400 Даташит, Описание, Даташиты
Classic Line 4-Cl2-200 Sensor
Temperature Dependence
4BU
Output vs Temperature
Average
140
+95%conf.
120
100
80
60
40
-20 -10
0
10 20
Temperature (0C)
-95%conf.
30 40
50
Baseline vs Temperature
3
2
1
0
-40 -20
0
-1
-2
-3
20 40
average
+ 95% conf
- 95% conf
Temperature (0C)
60
Cross-sensitivity Data
Gas
Hydrogen Sulfide
Carbon Monoxide
Sulphur Dioxide
Nitric Oxide
Nitrogen Dioxide
Hydrogen
Ammonia
Carbon Diooxide
Chlorine Dioxide
Concentration (ppm)
20
100
20
35
10
3000
100
10000
1
Output Signal (ppm Cl2 equivalent)
-4
0
0
0
12
0
0
0
3.5
Notes:
1. Calibration with cross sensitivity gas is not recommended
2. The cross sensitivity may fluctuate between +/- 30% and may differ from batch to batch or from sensor’s life time
3. The cross sensitivities are including but not limited to the above gases . It may also respond to other gases.
Solidsense GmbH - Felix-Wankel-Str. 5 - 82152 Krailling, Germany
Tel: +49 89 893 255 21 – Fax: +49 89 850 9374 – [email protected] – www.solidsense.de
CLE 4-Cl2-200 rev120727.doc - Page 2 of 2










Скачать PDF:

[ CLE-0922-400.PDF Даташит ]

Номер в каталогеОписаниеПроизводители
CLE-0922-400Chlorine SensorSolidsense
Solidsense

Номер в каталоге Описание Производители
TL431

100 мА, регулируемый прецизионный шунтирующий регулятор

Unisonic Technologies
Unisonic Technologies
IRF840

8 А, 500 В, N-канальный МОП-транзистор

Vishay
Vishay
LM317

Линейный стабилизатор напряжения, 1,5 А

STMicroelectronics
STMicroelectronics

DataSheet26.com    |    2020    |

  Контакты    |    Поиск